Esco » PCR Cabinets » Products: Streamline® Polymerase Chain Reaction Cabinets

Streamline® Polymerase Chain Reaction Cabinets

Streamline® Polymerase Chain Reaction Cabinets Picture
Bookmark and Share

Polymerase Chain Reaction* (PCR) is a process where millions of copies of DNA are amplified from a single copy, or low copy number template. This reaction is fundamental to almost all applications requiring high copy number of starting material and is used in all laboratories working with DNA and RNA. Because of the high copy number that is generated during PCR, it is absolutely essential to prevent contamination of the PCR reaction.

In addition to good laboratory practice, the ideal PCR laboratory should consist of three areas, each isolated from the other. Reagents should be prepared in the reagent preparation area and transferred to the sample preparation area, through a pass box, or inside closed containers. After preparation of the final reaction mix, the tubes should be transferred to the amplification area, again through a pass box or in a closed container. The PCR amplification and results analysis takes place in this area. To guarantee contaminant-free samples, it is essential to work in an environment where the air quality is controlled, such as a PCR clean bench, vertical laminar flow or biological safety cabinet.


Key Benefits:
  • HEPA-filtered laminar flow 
  • Cost & space saving
  • Simple & Reliable operation

 

ISOCIDE

  • True vertical laminar airflow with a large filter covering the entire work zone within the cabinet for better protection against cross contamination, as compared to conventional dead-air boxes.
  • High-quality polyurethane pre-filter and main HEPA filter with a typical efficiency of >99.99% at 0.3 microns provide the best product protection in the world; typical HEPA filter lifespan is more than 3 years depending on ambient operating conditions and total number of hours in usage.
  • A powerful 253.7 nanometer 20-watt UV lamp is placed behind the front panel out of the operator's direct line of sight, ensuring maximum operator comfort and safety; This unique placement design eliminates "dead zones" and ensures all exposed interior surfaces are decontaminated effectively.
  • Easily accessible adjustable UV-timer (from 0 - 60 minutes) for precise decontamination cycle control.
  • Integral safety cover (constructed of 5mm / 0.2" UV-absorbing beta radiation resistant polycarbonate - superior to conventional materials such as acrylic) with two operating positions; fully open (self-supporting with spring hinges) during normal use & fully closed during UV decontamination of interiors.
  • Effective safety features to prevent operator exposure to harmful UV rays; interlock system ensures UV lamps do not activate unless safety cover is closed; in case the cover is accidentally opened the UV lamps will also automatically deactivate.
  • Built-in warm white, 5.000k electronically ballasted lighting offers excellent illumination throughout the work zone in order to reduce operator fatigue and is comfortable to the eyes. Light tubes are mounted out of the air stream for better airflow uniformity.
Model No. External
Dimensions (mm)
Internal
Dimensions (mm)
Airflow
Velocity (Inflow)
Airflow
Velocity (Downflow)
Electrical
SCR-2A2
730 x 617 x 950
630 x 538 x 550
-
0.50 - 0.60(m/s) /
100 - 120(fpm)
110-120V,AC, 60Hz,1ø
Streamline® Polymerase Chain Reaction Cabinets Airflow

  • During operation, room air is drawn through the top of the cabinet via a pre-filter, with 85% arrestance, trapping larger particles and increasing the life of the main filter.
  • The air is then forced evenly through the main HEPA filter resulting in a unidirectional stream of clean air that is projected vertically over the internal work zone. All airborne contaminants are flushed and diluted, resulting in a particulate-free work environment.
  • The purified air then leaves the main work chamber across the open front of the cabinet.
  • The average airflow velocity of 0.50-0.60 m/s (100-120 fpm) in the SCR-2A ensures that cleanliness is maintained in the work zone.

STANDARDS COMPLIANCE CABINET PERFORMANCE AIR QUALITY FILTRATION ELECTRICAL SAFETY
IEST-RP-CC002.2, Worldwide ISO14644.1 Class 3, Worldwide
AS1386 Class 1.5, Australia
JISB9920 Class 3, japan
IEST-RP-CC034.1, Worldwide
IEST-RP-CC007.1, Worldwide
IEST-RP-CC001.3, Worldwide
EN1822 (H13), Europe
IEC61010-1, Worldwide
EN61010-1, Europe
UL61010-1, USA
CAN/CSA-22.2, No.61010-1